Reduced graphene oxide-functionalized polymer microneedle for continuous and wide-range monitoring of lactate in interstitial fluid

Reduced graphene oxide-functionalized polymer microneedle for continuous and wide-range monitoring of lactate in interstitial fluid

Despite substantial developments in minimally invasive lactate monitoring microneedle electrodes, most such electrode developments have focused on either sensitivity or invasiveness while ignoring a wide range of detection, which is the most important factor in measuring the normal range of lactate in interstitial fluid (ISF). Herein, we present a polymer-based planar microneedle electrode fabrication using microelectromechanical and femtosecond laser technology for the continuous monitoring of lactate in ISF. The microneedle is functionalized with two-dimensional reduced graphene oxide (rGO) and electrochemically synthesized platinum nanoparticles (PtNPs). A particular quantity of Nafion (1.25 wt%) is applied on top of the lactate enzyme to create a diffusion-controlled membrane. Due to the combined effects of the planar structure of the microneedle, rGO, and membrane, the biosensor exhibited excellent linearity up to 10 mM lactate with a limit of detection of 2.04 μM, high sensitivity of 43.96 μA mM−1cm−2, a reaction time of 8 s and outstanding stability, selectivity, and repeatability. The feasibility of the microneedle is evaluated by using it to measure lactate concentrations in artificial ISF and human serum. The results demonstrate that the microneedle described here has great potential for use in real-time lactate monitoring for use in sports medicine and treatment.